Power Consumption | 0.4W | Power Supply | ±12V |
Output Impedance | 50Ω | Operating Temperature | 0°C to 40°C (for standard products) -20°C to +40°C (for custom products) |
Energy Resolution | <2% @ 137Cs, 662 keV | Energy Range | 10 keV (integrated with Be windows) to 3 MeV |
Integral Nonlinearity | 10 keV to 1.3 MeV <0.5%; 10 keV to 3 MeV <1.0% |
Operating Environment | Non-condensing: 0°C to 40°C (<4% humidity) Adjustable conditions: -20°C to +0°C (<5% humidity) |
High Voltage Input | DC+800V@5x5x2.5mm CZT DC+900V@10x10x5mm CZT |
Crystal | 10x10x5mm/5x5x2.5mm Hemispherical CZT |
Signal Output | Exponential Decay Signal, Tr<150ns | Connection | 5-Core Connector |
Peak Drift Under Different Temperature | <4%@0°C~40°C <5%@-20°C~0°C | Peak Drift over 8h | <1% |
Size | ϕ30X57 mm | Weight | ≤60g |
Cadmium Zinc Telluride, also known as CdZnTe, abbreviated as CZT, is a semiconductor crystal. Its unique combination of characteristics including high energy resolution attributed to its large atomic number, and being able to operate at room temperature without cooling makes it a go-to option for applications where high sensitives and accurate imaging are essential (e.g. medical diagnostics.). In addition, in contrast with the NaI scintillation detectors that integrate PMT, CZT detectors transform incident photons into digital signals directly.
The CZTDT-01C1 high-energy resolution CZT (CdZnTe) Hemispherical Detectors represent a cutting-edge solution in radiation detection, leveraging the advanced capabilities of hemispherical Cadmium Zinc Telluride (CZT) crystals. This compact and efficient high-energy CZT detector from Shalom EO is engineered for the precise capture and analysis of the energy spectrum of high-energy photons, like gamma radiation and x radiation, ensuring exceptional performance across various applications with high energy resolution.
Shalom EO's CZTDT-01C1 high-energy resolution CZT (CdZnTe) Hemispherical Detectors' innovative design minimizes power consumption while enhancing stability, making it ideal for sensitive radiation detection tasks. Consisting of CZT crystals (copious dimensions accessible for selection) and low noise charge-sensitive preamplifier circuits, this high-energy resolution detector can accommodate the rigorous demands of professionals across multiple sectors, ensuring reliable and accurate radiation detection.
Application Areas:
Graphs:
The CZTDT-01C1 detector highlights outstanding spectral performance, precisely discriminating between different radiation sources. Its energy spectrum analysis capabilities include isotopes such as 241Am, 57Co, 137Cs, and 232Th.
The graphs below demonstrate the outstanding performance of Shalom EO's CZTDT-01C1 high-energy resolution CZT (CdZnTe) Hemispherical Detectors to deliver clear and high-resolution detection results across a wide range of energy levels.
1. The energy spectrum of CZTDT-01C1 CZT (CdZnTe) Hemispherical Detectors under 232TH
2. The energy spectrum of CZTDT-01C1 CZT (CdZnTe) Hemispherical Detectors under 241Am, 57Co, 137Cs